ObjectiveTo determine the agreement of high definition oscillometry (HDO) with direct arterial blood pressure measurements in normotensive, hypotensive and hypertensive horses during general anaesthesia.Study designExperimental study.AnimalsSeven healthy warmblood horses, aged 3–11 years, weighing 470–565 kg.MethodsMeasurements from a HDO device with the cuff placed around the base of the tail were compared with pressures measured invasively from the facial artery. High blood pressures were induced by intravenous (IV) administration of dobutamine (5 μg kg−1 minute−1) over ten minutes followed by norepinephrine (0.1 mg kg−1 IV) and low pressures by increasing the inspired fraction of isoflurane and administration of nitroglycerine (0.05 mg kg−1 IV). For analysis three pressure levels were determined: high (MAP>110 mmHg), normal (60 mmHgResultsA total of 245 paired measurements of systolic (SAP), mean (MAP) and diastolic (DAP) pressures were obtained. The HDO device underestimated blood pressure at hypertensive and normotensive levels and overestimated blood pressure at hypotensive levels. Best agreement was obtained for SAP and MAP within normotensive limits. At normotension, bias ± standard deviation for SAP, MAP and DAP were 0.1 ± 19.4 mmHg, 0.5 ± 14.0, 4.7 ± 15.6, respectively. At high pressure levels bias and SD were 26.1 ± 37.3 (SAP), 4.2 ± 19.4 (MAP), 1.5 ± 16.8 (DAP) and at low pressures -20.0 ± 20.9 (SAP), -11.4 ± 19.6 (MAP), -4.7 ± 20.1 (DAP), with HDO measurements at a MAP <50 mmHg often failing.Conclusion and clinical relevanceGood agreement with invasive arterial blood pressures was obtained with HDO at normotensive levels in horses. At high and low pressure ranges HDO was unreliable. Therefore, if haemodynamic instability is expected, invasive measurement remains preferable.  相似文献   

2.
ObjectiveTo assess the agreement between an oscillometric device and invasive blood pressure (IBP) measurements in anesthetized healthy adult guinea pigs.Study designProspective experimental study.AnimalsA total of eight adult Hartley guinea pigs.MethodsAll animals were anesthetized; a carotid artery was surgically exposed and catheterized for IBP measurements. A size 1 cuff placed on the right thoracic limb was connected to an oscillometric device for noninvasive blood pressure (NIBP) assessment. Concurrent pairs of systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressures were recorded simultaneously with both methods every 3 minutes for 30 minutes. Agreement between IBP and NIBP measurements was determined using the Bland–Altman method, considering the recommended standards for the validation of NIBP measurement devices proposed by the American College of Veterinary Internal Medicine (ACVIM).ResultsThe bias and the 95% limits of agreement were: –14 (–31 to 3) mmHg, –2 (–14 to 10) mmHg and –1 (–13 to 11) mmHg for SAP, DAP and MAP, respectively.Conclusions and clinical relevanceThe oscillometric device used in this study to measure NIBP did not meet ACVIM criteria for validation. It showed good agreement for DAP and MAP but not for SAP measurements. Considering the small size of these animals and the resulting difficulty in performing percutaneous arterial catheterization, this device might be a useful tool to assess MAP and DAP during anesthetic procedures in adult guinea pigs.  相似文献   

3.
ObjectiveTo evaluate the agreement between invasive blood pressure (IBP) and Doppler ultrasound blood pressure (DUBP) using three cuff positions and oscillometric blood pressure (OBP) in anesthetized dogs.Study designProspective study.AnimalsNine adult dogs weighing 14.5–29.5 kg.MethodsThe cuff was placed above and below the tarsus, and above the carpus with the DUBP and above the carpus with the OBP monitor. Based on IBP recorded via a dorsal pedal artery catheter, conditions of low, normal, and high systolic arterial pressures [SAP (mmHg) <90, between 90 and 140, and >140, respectively] were induced by changes in isoflurane concentrations and/or dopamine administration. Mean biases ± 2 SD (limits of agreement) were determined.ResultsAt high blood pressures, regardless of cuff position, SAP determinations with the DUBP underestimated invasive SAP values by more than 20 mmHg in most instances. With the DUBP, cuff placement above the tarsus yielded better agreement with invasive SAP during low blood pressures (0.2 ± 16 mmHg). The OBP underestimated SAP during high blood pressures (?42 ± 42 mmHg) and yielded better agreement with IBP for mean (MAP) and diastolic (DAP) arterial pressure measurements [overall bias: 2 ± 15 mmHg (MAP) and 0.2 ± 16 mmHg (DAP)].ConclusionsAgreement of SAP determinations with the DUBP is poor at SAP > 140 mmHg, regardless of cuff placement. Measurement error of the DUBP with the cuff placed above the tarsus is clinically acceptable during low blood pressures. Agreement of MAP and DAP measurements with this OBP monitor compared with IBP was clinically acceptable over a wide pressure range.Clinical relevanceWith the DUBP device, placing the cuff above the tarsus allows reasonable agreement with IBP obtained via dorsal pedal artery catheterization. Only MAP and DAP provide reasonable estimates of direct blood pressure with the OBP monitor evaluated.  相似文献   

4.
Objective To use the American College of Veterinary Internal Medicine (ACVIM) validation criteria to evaluate the performance of high definition oscillometric (HDO) and Doppler blood pressure measurement techniques against invasive blood pressure measurements in anaesthetized dogs. Study design Prospective clinical study. Animals Twenty client‐owned dogs. Materials and Methods Invasive blood pressure was measured using a catheter inserted into a pedal artery and an electronic transducer. The sites of cuff placement for the HDO measurements were the mid antebrachium or the proximal tail and, for the Doppler technique, the distal tibia. Agreement between invasive and non‐invasive blood pressure measurements was estimated by the Bland–Altman method. Results Only 10% and 34% of Doppler measurements were within 10 and 20 mmHg of invasive blood pressure values, respectively. The Doppler device failed to meet the ACVIM validation criteria for blood pressure measurement devices. The best agreement between HDO and invasive blood pressure measurement technique was observed for mean arterial blood pressure (MAP); 67% and 95% of readings were within 10 and 20 mmHg of invasive blood pressure values respectively. In addition, 52% and 87% of diastolic arterial blood pressure (DAP) measurements were within 10 and 20 mmHg of invasive readings. High definition oscillometric readings did not meet ACVIM recommended limits for SAP. Conclusion and clinical relevance The Doppler technique overestimated and the HDO device showed limited agreement with invasive blood pressure measurement in anaesthetized dogs. High definition oscillometry met most of the ACVIM requirements for MAP and DAP while the Doppler technique did not.  相似文献   

5.
ObjectiveTo assess agreement between carotid arterial pressure and auricular arterial, thoracic limb Doppler or thoracic limb oscillometric blood pressure measurements.Study designProspective experimental study.AnimalsSix adult New Zealand white rabbits.MethodsRabbits were anesthetized with isoflurane in oxygen at 1, 1.5 and 2 MAC on two separate occasions. Catheters in the auricular and the contralateral external carotid artery were connected to calibrated pressure transducers via non-compliant tubing. Inflatable cuffs of width equal to approximately 40% of the limb circumference were placed above the carpus on both thoracic limbs with a Doppler transducer placed distal to the cuff on one. Systolic (SAP) and mean (MAP) arterial blood pressure measurements were obtained at each dose, on each occasion. Agreement between measurement techniques was evaluated by repeated measures Bland Altman analysis with carotid pressure as the reference. Variation in bias over the measurement range was evaluated by regression analysis.ResultsCarotid MAP and SAP ranged from 20 to 65 mmHg and 37 to 103 mmHg respectively. Bias and 95% limits of agreement for auricular and oscillometric MAP were 7 (0–14) and ?5 (?21–11) mmHg, respectively, and for auricular, oscillometric and Doppler SAP were 23 (8–37), ?2 (?24–20) and 13 (?14–39) mmHg, respectively. Bias varied significantly over the measurement range (p < 0.001) for all three SAP techniques but not for MAP measurements.Conclusions and clinical relevanceLimits of agreement for all measurements were large but less so for MAP than SAP. Variation in bias with SAP should be considered when using these measurements clinically.  相似文献   

6.
ObjectiveTo use American College of Veterinary Internal Medicine (ACVIM) criteria to evaluate a high-definition oscillometric (HDO) blood pressure monitoring device versus invasive blood pressure (IBP) measurement in normotensive rabbits anaesthetized with two different anaesthetic protocols.Study designProspective experimental study.AnimalsA group of 20 healthy adult New Zealand White rabbits weighing 4.36 ± 0.37 kg (mean ± standard deviation).Materials and methodsRabbits were premedicated with butorphanol 0.5 mg kg–1 and midazolam 0.5 mg kg–1 subcutaneously (SC, group BMA) or ketamine 25 mg kg–1 and medetomidine 0.4 mg kg–1 SC (group KM). Anaesthesia was induced with alfaxalone administered intravenously (group BMA) or isoflurane by face mask (group KM) and maintained with isoflurane in oxygen. IBP was measured from the central auricular artery. The cuff for the HDO monitor was placed distal to the left elbow and distal to the left tarsus. Agreement between invasive and HDO measurements was evaluated using Bland–Altman method.ResultsIn group KM there was better agreement between the HDO device and IBP when the cuff was placed on the thoracic limb, with 100% and 91% of the readings for mean (MAP) and diastolic arterial pressure (DAP), respectively, within 10 mmHg of the IBP measurements. The agreement, although worse, also met the ACVIM criteria for systolic arterial pressure (SAP; 53% of the readings within 10 mmHg). In group BMA, the device met the criteria with the cuff on the thoracic limb only, and only for MAP and DAP (73% and 75% of the measurements within 10 mmHg of the IBP, respectively) but not for SAP (12%).Conclusion and clinical relevanceThe HDO device met most of the ACVIM criteria for noninvasive blood pressure measurement in anaesthetized rabbits, specifically when the cuff was placed distal to the elbow and the anaesthetic protocol included ketamine and medetomidine.  相似文献   

7.
ObjectiveTo assess agreement between oscillometric noninvasive blood pressure (NIBP) measurements using LifeWindow monitors (LW9xVet and LW6000V) and invasive blood pressure (IBP). To assess the agreement of NIBP readings using a ratio of cuff width to mid-cannon circumference of 25% and 40%.Study designProspective, randomized clinical study.AnimalsA total of 43 adult horses undergoing general anesthesia in dorsal recumbency for different procedures.MethodsAnesthetic protocols varied according to clinician preference. IBP measurement was achieved after cannulation of the facial artery and connection to an appropriately positioned transducer connected to one of two LifeWindow multiparameter monitors (models: LW6000V and LW9xVet). Accuracy of monitors was checked daily using a mercury manometer. For each horse, NIBP was measured with two cuff widths (corresponding to 25% or 40% of mid-cannon bone circumference), both connected to the same monitor, and six paired IBP/NIBP readings were recorded (at least 3 minutes between readings). NIBP values were corrected to the relative level of the xiphoid process. A Bland–Altman analysis for repeated measures was used to assess bias (NIBP–IBP) and limits of agreement (LOAs).ResultsThe 40% cuff width systolic arterial pressure [SAP; bias 7.9 mmHg, LOA –26.6 to 42.3; mean arterial pressure (MAP): bias 4.9 mmHg, LOA –28.2 to 38.0; diastolic arterial pressure (DAP): bias 4.2 mmHg, LOA –31.4 to 39.7)] performed better than the 25% cuff width (SAP: bias 26.4 mmHg, LOA –21.0 to 73.9; MAP: bias 15.7 mmHg, LOA –23.8 to 55.2; DAP: bias 10.9 mmHg, LOA –33.2 to 54.9).Conclusions and clinical relevanceUsing the LifeWindow multiparameter monitor in anesthetized horses, the 40% cuff width provided better agreement with IBP; however, both cuff sizes and both monitor models failed to meet American College of Veterinary Internal Medicine Consensus Statement Guidelines.  相似文献   

8.
OBJECTIVE: To compare the performance of the Surgivet Non-Invasive Blood Pressure (NIBP) monitor V60046 with an invasive blood pressure (IBP) technique in anaesthetized dogs. STUDY DESIGN: A prospective study. ANIMALS: Thirty-four dogs, anaesthetized for a variety of procedures. METHODS: Various anaesthetic protocols were used. Invasive blood pressure measurement was made using a catheter in the femoral or the pedal artery. A cuff was placed on the contralateral limb to allow non invasive measurements. Recordings of arterial blood pressures (ABPs) were taken at simultaneous times for a range of pressures. For analysis, three pressure levels were determined: high [systolic blood pressure (SAP) > 121 mmHg], normal (91 mmHg < SAP < 120 mmHg) and low (SAP < 90 mmHg). Comparisons between invasive and non invasive measurements were made using Bland-Altmann analysis. RESULTS: The NIBP monitor consistently underestimated blood pressure at all levels. The lowest biases and greatest precision were obtained at low and normal pressure levels for SAP and mean arterial pressure (MAP). At low blood pressure levels, the biases +/- 95% confidence interval (CI) were 1.9 +/- 2.96 mmHg (SAP), 8.3 +/- 2.41 mmHg diastolic arterial pressure (DAP) and 3.5 +/- 2.09 mmHg (MAP). At normal blood pressure levels, biases and CI were: 1.2 +/- 2.13 mmHg (SAP), 5.2 +/- 2.32 mmHg (DAP) and 2.1 +/- 1.54 mmHg (MAP). At high blood pressure levels, the biases and CI were 22.7 +/- 5.85 mmHg (SAP), 5.5 +/- 3.13 mmHg (DAP) and 9.4 +/- 3.52 mmHg (MAP). In 90.6% of cases of hypotension (MAP < 70 mmHg), the low blood pressure was correctly diagnosed by the Surgivet. CONCLUSIONS: Measurement of blood pressure with the indirect monitor allowed detection of hypotension using either SAP or MAP. The most accurate readings were determined for MAP at hypotensive and normal levels. The monitor lacked accuracy at high pressures. CLINICAL RELEVANCE: When severe challenges to the cardiovascular system are anticipated, an invasive method of recording ABP is preferable. For routine usage, the Surgivet monitor provided a reliable and safe method of NIBP monitoring in dogs, thereby contributing to the safety of anaesthesia by providing accurate information about the circulation.  相似文献   

9.
ObjectiveTo determine the accuracy of an oscillometric blood pressure monitor in anesthetized sheep.Study designProspective study.AnimalsTwenty healthy adult sheep, 11 males and nine females, weighing 63.6 ± 8.6 kg.MethodsAfter premedication with buprenorphine or transdermal fentanyl, anesthesia was induced with ketamine‐midazolam and maintained with isoflurane and ketamine, 1.2 mg kg?1 hour?1, ± lidocaine, 3 mg kg?1 hour?1. Invasive blood pressure measurements were obtained from an auricular arterial catheter and noninvasive measurements were from a cuff on the metatarsus or antebrachium. Simultaneous invasive and noninvasive measurements were recorded over a range (55–111 mmHg) of mean arterial pressures (MAP). Isoflurane concentration was increased to decrease MAP and decreasing the isoflurane concentration and infusing dobutamine achieved higher pressures. Invasive and noninvasive measurements were compared.ResultsCorrelation (R2) was good between the two methods of measurement (average of three consecutive readings) for systolic (SAP) (0.87), diastolic (DAP) (0.86), and mean (0.90) arterial pressures (p < 0.001). Bias ± SD between noninvasive and invasive measurements for SAP was 3 ± 8 mmHg, for DAP was ?10 ± 7 mmHg, and MAP was ?7 ± 6 mmHg. There was no significant difference between the average of three measurements and use of the first measurement. Correlations using the first measurement were SAP (0.82), DAP (0.84), and MAP (0.89). Bias ± SD for SAP was 3 ±10 mmHg, for DAP was ?11 ± 7 mmHg, and MAP was ?7 ± 6 mmHg. The oscillometric monitor slightly overestimated SAP and underestimated DAP and MAP for both average values and the first reading.Conclusions and clinical relevanceThis oscillometric model provided MAP measurements that were acceptable by ACVIM standards. MAP measurements with this monitor were lower than those found with the invasive technique so a clinical diagnosis of hypotension may be made in sheep that are not hypotensive.  相似文献   

10.
ObjectiveTo evaluate agreement with central systemic arterial pressure of an oscillometer and two cuff widths placed on the thoracic or pelvic limbs.Study designProspective experimental study.AnimalsA group of nine New Zealand White rabbits weighing 3.5 ± 0.3 kg.MethodsRabbits were sedated with dexmedetomidine and midazolam, then anesthetized with ketamine and sevoflurane. The femoral artery was surgically exposed and a 20 gauge, 5 cm catheter inserted to measure systolic (SAP), mean (MAP) and diastolic (DAP) blood pressure at the iliac artery and caudal aorta junction. Adjustments of vaporizer dial and dobutamine infusion provided a range of invasive blood pressure (IBP). Two measurements of IBP were recorded during the oscillometer cycling phase, and the mean value was used in analyses. Oscillometer cuffs of bladder width 2.0 cm (S1) and 2.5 cm (S2) were placed proximal to the carpus and tarsus. Cuff width to circumference ratio was calculated. Oscillometer SAP, MAP and DAP were paired with corresponding IBP values. Agreement was assessed using linear mixed models (p < 0.05).ResultsCuff ratios for both limbs were 41% (S1 cuff) and 50% (S2 cuff) and 122–139 paired observations were obtained. There was significant limb × cuff interaction with SAP and MAP. The oscillometer overestimated SAP and MAP on the pelvic limb and underestimated SAP and MAP on the thoracic limb. For SAP, the oscillometer overestimated by constant bias (–19 ± 2 mmHg) and proportional bias (0.28 ± 0.02 mmHg per 1 mmHg increase). For MAP, the oscillometer underestimated by constant bias (4 ± 2 mmHg) and was worse with S2 on the thoracic limb. Overestimation was similar between cuffs on the pelvic limb. For DAP, the oscillometer underestimated by constant bias (15 ± 2 mmHg).Conclusions and clinical relevanceCuff S1 on the thoracic limb provided best estimation of MAP.  相似文献   

11.
ObjectiveTo evaluate the agreement of two noninvasive blood pressure devices: a human device with the cuff placed on the wrist (Omron R1) and a veterinary device with the cuff placed on the upper brachium (Surgivet Advisor Vital Signs Monitor) with invasive blood pressure (IBP) measurement in anaesthetized chimpanzees.Study designProspective clinical study.AnimalsA convenience sample of 11 adult chimpanzees undergoing anaesthesia for translocation and routine health checks.MethodsSystolic (SAP) and diastolic arterial pressures (DAP) were continuously recorded via a transducer connected to a femoral artery cannula, and at 5 minute intervals from the two oscillometric devices. Agreement was explored using Bland-Altman analysis and bias defined as the mean difference between the two measurement methods. Spearman correlation coefficients were calculated. Significance was set at p < 0.05.ResultsBias and standard deviation for the Surgivet compared with IBP were 8.6 ± 18 for SAP and 8.4 ± 9.9 for DAP, showing a significant underestimation of both variables. Limits of agreement (LOA) were from –27 to 44 for SAP and from –11 to 28 for DAP. Correlation coefficients between the Surgivet and IBP values were 0.86 for SAP and 0.85 for DAP (p < 0.0001). Bias and standard deviation for the Omron compared with the IBP were –21 ± 25 for SAP and –18 ± 15 for DAP, showing a significant overestimation of both variables. LOA were from –70 to –28 for SAP and from –47 to 11 for DAP. Spearman correlation coefficients between the Omron and IBP values were 0.64 for SAP and 0.72 for DAP (p < 0.0001).Conclusions and clinical relevanceAlthough neither device met all the criteria for device validation, the Surgivet presented better agreement with IBP values than the Omron in adult anaesthetized chimpanzees.  相似文献   

12.
ObjectiveTo compare Doppler and oscillometric methods of indirect arterial blood pressure (IBP) with direct arterial measurements in anesthetized and awake red-tailed hawks.Study designProspective, randomized, blinded study.AnimalsSix, sex unknown, adult red-tailed hawks.MethodsBirds were anesthetized and IBP measurements were obtained by oscillometry (IBP-O) and Doppler (IBP-D) on the pectoral and pelvic limbs using three cuffs of different width based on limb circumference: cuff 1 (20–30% of circumference), cuff 2 (30–40%), and cuff 3 (40–50%). Direct arterial pressure measurements were obtained from the contralateral superficial ulnar artery. Indirect blood pressure measurements were compared to direct systolic arterial pressure (SAP) and mean arterial pressure (MAP) during normotension and induced states of hypotension and hypertension. Measurements were also obtained in awake, restrained birds. Three-way anova, linear regression and Bland–Altman analyses were used to evaluate the IBP-D data. Results are reported as mean bias (95% confidence intervals).ResultsThe IBP-O monitor reported errors during 54% of the measurements. Indirect blood pressure Doppler measurements were most accurate with cuff 3 and were comparable to MAP with a bias of 2 (?9, 13 mmHg). However, this cuff consistently underestimated SAP with a bias of 33 (19, 48 mmHg). Variability in the readings within and among birds was high. There was no significant difference between sites of cuff placement. Awake birds had SAP, MAP and diastolic arterial pressure that were 56, 43, and 38 mmHg higher than anesthetized birds.Conclusions and clinical relevanceIndirect blood pressure (oscillometric) measurements were unreliable in red-tailed hawks. Indirect blood pressure (Doppler) measurements were closer to MAP measurements than SAP measurements. There was slightly better agreement with the use of cuff 3 on either the pectoral or pelvic limbs. Awake, restrained birds have significantly higher arterial pressures than those under sevoflurane anesthesia.  相似文献   

13.
ObjectiveTo evaluate the agreement between oscillometric blood pressure (OBP) measured from the tongue and invasive blood pressure (IBP), and to compare OBPs measured from the tongue with OBPs measured from the pelvic limb and tail.Study designProspective experimental study.AnimalsA total of eight adult Beagle dogs weighing 11.1 ± 1.2 kg.MethodsAnimals were premedicated with intravenous (IV) acepromazine (0.005 mg kg–1). Anesthesia was induced with alfaxalone (3 mg kg–1) IV and maintained with isoflurane. The dorsal pedal artery was catheterized for IBP measurements. Systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressure were simultaneously measured from the tongue, pelvic limb and tail. Based on invasive SAP, hypertension (>140 mmHg), normotension (90–140 mmHg) and hypotension (<90 mmHg) were induced by controlling end-tidal isoflurane concentrations and/or dobutamine/dopamine administration. Agreement between paired IBP and OBP measurements was analyzed with reference standards for noninvasive blood pressure devices used in small animals and humans.ResultsRegardless of cuff placement, the mean bias ± standard deviation between IBP and OBP met veterinary (≤10 ± 15 mmHg) and human (<5 ± 8 mmHg) standards for MAP and DAP. SAP measurements provided by the OBP device showed unacceptable agreement with IBP, and the bias between methods increased at higher blood pressures, regardless of cuff site. During hypotension, tongue OBP showed the largest percentage of absolute difference <10 mmHg in relation to IBP for SAP (90%), MAP (97%), and DAP (93%), compared with pelvic limb (60%, 97% and 82%, respectively) and tail OBP (54%, 92% and 77%, respectively).Conclusions and clinical relevanceThe tongue is a clinically useful site for measuring OBP in anesthetized Beagle dogs, providing reliable estimates of MAP and DAP. The tongue could replace other cuff placement sites and may be a relatively suitable site for assessing hypotension.  相似文献   

14.
ObjectiveTo evaluate a veterinary-specific oscillometric noninvasive blood pressure (NIBP) system according to the guidelines of the American College of Veterinary Internal Medicine (ACVIM) Consensus Statement.Study designProspective clinical study.AnimalsA total of 33 client-owned cats (20 females and 13 males).MethodsCats were premedicated with methadone (0.3 mg kg−1) and alfaxalone (2 mg kg−1) intramuscularly. After 15 minutes anesthesia was induced with isoflurane (3%) in 100% oxygen by facemask while breathing spontaneously. A 22 gauge catheter was placed in the median caudal artery and systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressures were measured. NIBP measurements were collected by placing the cuff (40% of limb circumference) on the right or left antebrachium. The agreement between the two methods was evaluated with the Bland–Altman methods, and the oscillometric NIBP device was evaluated using the ACVIM guidelines for validation of devices.ResultsData from 30 of the 33 cats were analyzed. Five paired measurements were taken from each cat, totaling 150 paired measurements. Mean bias (limits of agreements) for SAP, DAP and MAP were 2.7 (−22.7 to 28.1), 0.9 (−22.3 to 24.2) and 1.3 (−20.4 to 23.0). The oscillometric NIBP passed all validation criteria, except correlation which was <0.9 for SAP, DAP and MAP.Conclusions and clinical relevanceThe Vet20 did not meet all validation criteria by the ACVIM. However, all criteria except correlation were met.  相似文献   

15.

Objective

To determine agreement between invasive blood pressures measured in three peripheral arteries in anaesthetized horses undergoing elective surgery.

Study design

Prospective balanced incomplete block design.

Animals

A total of 18 client-owned horses.

Methods

Invasive blood pressure (IBP) was measured simultaneously in one of the following three combinations: 1) transverse facial and facial artery; 2) transverse facial and metatarsal artery; and 3) facial and metatarsal artery. The agreement in blood pressure measured for each combination was performed in six horses. At each sample time, systolic (SAP), mean (MAP) and diastolic (DAP) arterial pressures were measured concurrently in each artery, and the mean of three consecutive measurements was recorded. The position of horse, heart rate and use of dobutamine were also recorded. Bland–Altman analysis was used to assess agreement between sites.

Results

A total of 54 paired measurements were obtained, with 18 paired measurements from each combination. All paired measurements showed poor and haphazard (nonsystematic) agreement. The widest limit of agreement was 51 mmHg for SAP measured in the facial artery and metatarsal artery, with a bias of –11 mmHg. The smallest limit of agreement was 16 mmHg for MAP measured in the transverse facial and metatarsal artery, with a bias of 1 mmHg.

Conclusions and clinical relevance

There was poor and haphazard agreement for SAP, MAP and DAP measured in each pair of peripheral arteries in this study. These results show that blood pressure measured in different peripheral arteries cannot be used interchangeably. This has implications for studies that use IBP as an outcome variable and studies determining agreement between noninvasive blood pressure and IBP measurements in horses under general anaesthesia.  相似文献   

16.

Objective

To determine the agreement of invasive blood pressure measured in the facial, metatarsal and carotid arteries, and evaluate the effects of two haemodynamic conditions on agreement.

Study design

Prospective randomized study.

Animals

A group of eight horses aged 7 (4–23) years with a body weight of 493 ± 33 kg.

Methods

Horses were anaesthetized and positioned in dorsal recumbency. Invasive blood pressure was measured simultaneously via catheters placed in the facial, metatarsal and carotid arteries. Cardiovascular function and agreement between arteries was assessed before and during administration of phenylephrine and sodium nitroprusside. These were administered until carotid mean pressure (MAPc) increased or decreased from baseline (65 ± 5) to >90 or <50 mmHg, respectively. Data recorded at each sample time included systolic (SAP), mean (MAP) and diastolic (DAP) arterial pressures for carotid (c), facial (f) and metatarsal (m) arteries as well as cardiac output (Q˙t) and systemic vascular resistance (SVR). Bland–Altman analysis was used to assess agreement between peripheral and central sites, and regression analysis to determine influence of Q˙t and SVR.

Results

The largest difference was observed in SAPc and SAPm with a bias and limits of agreement (LOA) of 2 (?15 to 19) mmHg. The bias (LOA) for MAPc and MAPf was 2 (?4 to 9) mmHg and for MAPc and MAPm was 5 (?4 to 14) mmHg. The best agreement for DAP was seen between DAPc and DAPf with bias (LOA) of 1 (?3 to 5) mmHg. Regression analysis indicated marginal influence on agreement by Q˙t on MAPc and MAPf.

Conclusions and clinical relevance

MAP and DAP of the carotid artery were higher than those of the peripheral arteries, which may lead to overzealous treatment of hypotension, albeit maintaining central pressures. Q˙t and SVR did not largely influence the difference between sites.  相似文献   

17.

Objective

To assess the validity and reliability of Doppler ultrasonography (DOP) as compared with invasive arterial blood pressure measurements in anaesthetized dogs weighing less than 5 kg.

Study design

Controlled, prospective, clinical study.

Animals

A total of 41 privately owned dogs weighing less than 5 kg.

Methods

The dogs were anaesthetized, and an intra-arterial catheter was placed aseptically in the dorsal pedal artery of the pelvic limb to perform invasive blood pressure (IBP) measurement. The contralateral metatarsal surface of the foot was clipped in order to perform DOP. Both techniques were used to record blood pressure measurements every 5 minutes during surgical procedures. The blood pressure measurements were categorized into two groups: hypotensive [mean arterial pressure (MAP) < 60 mmHg] and normotensive (MAP between 60 and 120 mmHg). A linear mixed model was used to compare the DOP and IBP values. The results were evaluated according to the requirements of the American College of Veterinary Internal Medicine (ACVIM) guidelines for the validation of devices.

Results

DOP provided higher values compared to the systolic arterial blood pressure (SAP) and MAP of IBP measurements. The closest agreement between the two techniques was found for SAP; the bias was 8.8, and limits of agreement (LOA) were –32.9 and 50.4. Similar results were observed when the IBP technique was categorized. The closest agreement was for SAP in animals categorized as normotensive; the bias was 8.2, and LOAs were –32.8 and 49.2. The level of agreement between DOP and IBP did not meet the ACVIM recommendations.

Conclusions and clinical relevance

Our results suggest there is poor agreement between DOP and IBP measurements in anaesthetized dogs weighing less than 5 kg. Hence, the use of DOP in these animals could be misleading.  相似文献   

18.
Oscillometric blood pressure monitoring may be a practical tool for short procedures or those performed outside of the operating room. Oscillometric and direct blood pressure values in 30 juvenile and adult horses in a clinical setting using mixed effect models were compared. The limits of agreement and percentage errors were also calculated. We evaluated the sensitivity and false positive rate for the oscillometric method to trigger an intervention for treating blood pressure [direct mean arterial pressure (MAP) < 70 mmHg]. Oscillometric MAP and diastolic arterial pressure (DAP) differed from direct values (P < 0.001); systolic arterial pressure (SAP) did not (P = 0.08). Wide limits of agreement were observed. Percentage errors were smaller for SAP (39%) than for MAP and DAP (48% and 72%). The oscillometric monitor indicated there was a requirement for blood pressure treatment with a true positive rate of 82%, consequently, it failed 18% of the times. The false positive rate (unnecessary treatment) was 55%.  相似文献   

19.
Objective – To determine the accuracy and precision of an oscillometric noninvasive blood pressure device as a predictor of invasive direct blood pressure in healthy anesthetized hypotensive and normotensive dogs. Design – Prospective observational study. Setting – University teaching hospital. Animals – Eight crossbred adult dogs. Interventions – Anesthesia was induced with propofol and maintained with isoflurane. A catheter was placed in the dorsal pedal artery to record systolic, mean, and diastolic arterial blood pressures (aSAP, aMAP, and aDAP, respectively). The noninvasive blood pressure device cuff was placed around the contralateral front limb to record noninvasive systolic, mean, and diastolic blood pressure (nSAP, nMAP, and nDAP). Two states of blood pressure (BP) were studied: baseline state was established by keeping end‐tidal isoflurane concentration at 1.2±0.1%. The hypotensive state was achieved by maintaining the same isoflurane concentration while withdrawing approximately 40% of the animal's blood volume until aMAP was stable at approximately 40 mm Hg. At the end of the study, blood was returned to the animal and it was allowed to recover from anesthesia. Measurements and Main Results – Agreement between the direct and indirect BP measurements was determined by the Bland‐Altman method. The SAP and MAP but not DAP bias varied significantly between each BP state. Normotensive absolute biases (mean [SD]) for SAP, MAP, and DAP were ?14.7 mm Hg (15.5 mm Hg), ?16.4 mm Hg (12.1 mm Hg), and ?14.1 mm Hg (15.8 mm Hg), respectively. Absolute biases during the hypotensive state for SAP, MAP, and DAP were ?32 mm Hg (22.6 mm Hg), ?24.2 mm Hg (19.5 mm Hg), and ?16.8 mm Hg (17.2 mm Hg), respectively. Conclusion – The oscillometric device was not reliably predictive of intra‐arterial BP during hypotension associated with acute hemorrhage.  相似文献   

20.
ObjectiveTo measure the level of agreement between Doppler measured (DOP) arterial blood pressure (ABP) in the forelimb and directly measured (DIR) auricular systolic ABP (SAP) and mean ABP (MAP) in isoflurane-anaesthetized rabbits.Study designProspective clinical study.AnimalsData were analysed from 17 of 24 healthy rabbits, weighing 1.3–2.8 kg.MethodsRabbits were anaesthetized for neutering using a standardized protocol. A 26G catheter placed in an auricular artery was connected via heparinised saline filled non-compliant tubing (regularly flushed) to a calibrated pressure transducer (zeroed level with the thoracic inlet) to obtain DIR ABP. A cuff was placed proximal to the carpus (approximately level with the thoracic inlet) and a Doppler transducer sited over the dorsal carpal branch of the radial artery to obtain DOP ABP. Simultaneous DIR and DOP ABP recordings were made every 5–10 minutes during anaesthesia. Agreement was assessed as described by Bland JM &; Altman (2007).ResultsMean ± SD cuff width: limb circumference ratio was 0.50 ± 0.04. Mean between-method differences ± SD, DIR SAP- DOP and DIR MAP- DOP, were +1 ± 8 and ?13 ± 8 mmHg respectively. The 95% limits of agreement between DIR SAP and DOP and between DIR MAP and DOP were ?14 to +17 and ?28 to +2 mmHg respectively. Differences between DIR SAP and DOP were ≤10 mmHg 85% of the time. Defining hypotension as either DIR SAP < 80 mmHg or DIR MAP < 60 mmHg, and taking DOP ABP of <80 mmHg to indicate hypotension, sensitivity and specificity were 92% and 67% respectively.ConclusionsGood agreement was found between DIR SAP and DOP. Doppler measurements below 80 mmHg are a reliable indicator of arterial hypotension.Clinical relevanceDOP is acceptable for monitoring ABP in isoflurane-anaesthetized rabbits and is useful for detection of hypotension.  相似文献   