Evaluation of the agreement of two oscillometric blood pressure devices with invasive blood pressure in anaesthetized chimpanzees (Pan troglodytes)
Institution:1. School of Veterinary Medicine and Science, University of Nottingham, Leicestershire, UK;2. Twycross Zoo, East Midland Zoological Society, Atherstone, UK
Abstract:ObjectiveTo evaluate the agreement of two noninvasive blood pressure devices: a human device with the cuff placed on the wrist (Omron R1) and a veterinary device with the cuff placed on the upper brachium (Surgivet Advisor Vital Signs Monitor) with invasive blood pressure (IBP) measurement in anaesthetized chimpanzees.Study designProspective clinical study.AnimalsA convenience sample of 11 adult chimpanzees undergoing anaesthesia for translocation and routine health checks.MethodsSystolic (SAP) and diastolic arterial pressures (DAP) were continuously recorded via a transducer connected to a femoral artery cannula, and at 5 minute intervals from the two oscillometric devices. Agreement was explored using Bland-Altman analysis and bias defined as the mean difference between the two measurement methods. Spearman correlation coefficients were calculated. Significance was set at p < 0.05.ResultsBias and standard deviation for the Surgivet compared with IBP were 8.6 ± 18 for SAP and 8.4 ± 9.9 for DAP, showing a significant underestimation of both variables. Limits of agreement (LOA) were from –27 to 44 for SAP and from –11 to 28 for DAP. Correlation coefficients between the Surgivet and IBP values were 0.86 for SAP and 0.85 for DAP (p < 0.0001). Bias and standard deviation for the Omron compared with the IBP were –21 ± 25 for SAP and –18 ± 15 for DAP, showing a significant overestimation of both variables. LOA were from –70 to –28 for SAP and from –47 to 11 for DAP. Spearman correlation coefficients between the Omron and IBP values were 0.64 for SAP and 0.72 for DAP (p < 0.0001).Conclusions and clinical relevanceAlthough neither device met all the criteria for device validation, the Surgivet presented better agreement with IBP values than the Omron in adult anaesthetized chimpanzees.
Keywords:anaesthesia  apes  invasive blood pressure  noninvasive blood pressure  primate  validation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号