首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of a Radioimmunoassay Technique Measuring Calcitonin in The Bovine,Ovine and Porcine Species
Authors:Kristina Forslund  Mats Stridsberg
Institution:Department of Pharmacology, Faculty of Veterinary Medicine, Swedish University of Agricultural Sciences, Biomedicum, Uppsala, Sweden
Abstract:A heterologous RIA system was set up for measuring bovine, ovine and porcine calcitonin (CT). The system consisted of porcine GT used as standard and for the preparation of an iodinated tracer. The antiserum used was raised against ovine GT. For each analysis was used 25–200 µl blood plasma. Practical detection limit was 0.25 µg of CT per litre of blood plasma.The parallelism between the dose response curves for the p-GT standard and for the assay of increasing amounts of bovine, ovine and porcine blood plasma showed the suitability of the present assay system to study the GT secretion in these species. Furthermore, the reliability of the method was verified by a clearly recognized CT response to calcium infusion.
Keywords:calcitonin  RIA  bovine  ovine  porcine  Ca infusion  linearization
设为首页 | 免责声明 | 关于勤云 | 加入收藏