首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of a combined transcutaneous carbon dioxide pressure and pulse oximetry sensor in adult sheep and dogs
Authors:Vogt Rainer  Rohling Roman  Kästner Sabine
Institution:Vetsuisse Faculty, Anesthesia Division, University of Zurich, Zurich, Switzerland.
Abstract:OBJECTIVE: To evaluate a combined transcutaneous carbon dioxide pressure (tcPCO(2)) and pulse oximetry sensor in sheep and dogs. ANIMALS: 13 adult sheep and 11 adult dogs. PROCEDURES: During inhalation anesthesia, for the first 10 minutes following sensor placement, arterial blood gas was analyzed and tcPCO(2) was recorded every 2 minutes. Subsequently, the animals were hyper-, normo-, and hypoventilated. The simultaneously obtained tcPCO(2) and PaCO(2) values were analyzed by use of Bland-Altman statistical analysis. RESULTS: Mean +/- SD overall difference between tcPCO(2) and PaCO(2) 10 minutes after sensor application was 13.3 +/- 8.4 mm Hg in sheep and 8.9 +/- 12 mm Hg in dogs. During hyper-, normo-, and hypoventilation, mean difference (bias) and precision (limits of agreement bias +/- 2 SD]) between tcPCO(2) and PaCO(2) values were 13.2 +/- 10.4 mm Hg (limits of agreement, -7.1 and 33.5 mm Hg) in sheep and 10.6 +/- 10.5 mm Hg (limits of agreement, -9.9 and 31.2 mm Hg) in dogs, respectively. Changes in PaCO(2) induced by different ventilation settings were detected by the tcPCO(2) sensor with a lag (response) time of 4.9 +/- 3.5 minutes for sheep and 6.2 +/- 3.6 minutes for dogs. CONCLUSIONS AND CLINICAL RELEVANCE: The tcPCO(2) sensor overestimated PaCO(2) in sheep and dogs and followed changes in PaCO(2) with a considerable lag time. The tcPCO(2) sensor might be useful for noninvasive monitoring of changes but cannot be used as a surrogate measure for PaCO(2).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号