Evaluation of a continuous glucose monitoring system in cats with diabetes mellitus |
| |
Authors: | Ristic Jelena M E Herrtage Michael E Walti-Lauger Sabine M M Slater Linda A Church David B Davison Lucy J Catchpole Briän |
| |
Institution: | Axiom Veterinary Laboratories, Teignmouth, Devon TQ14 8AH, UK. jelena.ristic@axiomvetlab.co.uk |
| |
Abstract: | A continuous glucose monitoring system (CGMS) was evaluated in 14 cats with naturally occurring diabetes mellitus. The device measures interstitial fluid glucose continuously, by means of a sensor placed in the subcutaneous tissue. All cats tolerated the device well and a trace was obtained on 15/16 occasions. There was good correlation between the CGMS values and blood glucose concentration measured using a glucometer (r=0.932, P<0.01). Limitations to the use of the CGMS are its working glucose range of 2.2-22.2 mmol/l (40-400 mg/dl) and the need for calibration with a blood glucose measurement at least every 12 h. When compared to a traditional blood glucose curve, the CGMS is minimally invasive, reduces the number of venepunctures necessary to assess the kinetics of insulin therapy in a patient and provides a truly continuous glucose curve. |
| |