Validation and standardization of virus neutralizing test using indirect immunoperoxidase technique for the quantification of antibodies to rabies virus |
| |
Authors: | Ogawa T Gamoh K Aoki H Kobayashi R Etoh M Senda M Hirayama N Nishimura M Shiraishi R Servat A Cliquet F |
| |
Affiliation: | National Veterinary Assay Laboratory, Ministry of Agriculture, Forestry and Fisheries, 1-15-1 Tokura, Kokubunji, Tokyo, Japan. ogawa@nval.go.jp |
| |
Abstract: | A virus neutralizing test using an indirect immunoperoxidase technique (VNT-IIP) for rabies has been developed for the titration of dog and cat serum samples in Japan. The VNT-IIP has the advantage that results obtained can be viewed by the naked eye. The purpose of this study was to validate the VNT-IIP and compare it with one of the international standard methods, the fluorescent antibody virus neutralization test (FAVNT). The VNT-IIP showed satisfactory repeatability, high analytical specificity and good accuracy. Regarding the comparison between the VNT-IIP and the FAVNT, the VNT-IIP showed good agreement (91.9%), high sensitivity (92.8%) as well as specificity (87.0%) and good correlation (r = 0.92). As described above, the validation of the VNT-IIP was satisfactory and the performances of the test proved to be equivalent to those of an international standard method. |
| |
Keywords: | Rabies virus neutralization test indirect immunoperoxidase technique fluorescent antibody virus neutralization test |
本文献已被 PubMed 等数据库收录! |
|