In vitro validation of a Pitot-based flow meter for the measurement of respiratory volume and flow in large animal anaesthesia
Authors:Yves PS Moens DVM,PD,PhD,Diplomate ECVAA,,Peter Gootjes&dagger   Ing,,Jean-Claude Ionita&Dagger   DVM,,Erkki Heinonen§   Lic Tech,PhD, &   Urs Schatzmann&Dagger   DVM,PhD,Diplomate ECVAA
Affiliation:Clinic of Anaesthesiology and Perioperative Intensive Care, University of Veterinary Medicine, Vienna, Austria;
Faculty of Veterinary Medicine, Utrecht University, Utrecht, the Netherlands;
Division of Anaesthesiology, Department of Clinical Veterinary Medicine, University of Bern, Bern, Switzerland;
Datex-Ohmeda Research Unit, Helsinki, Finland
Abstract:Objective  To remodel and validate commercially available monitors and their Pitot tube-based flow sensors for use in large animals, using in vitro techniques.
Study design  Prospective, in vitro experiment.
Methods  Both the original and the remodelled sensor were studied with a reference flow generator. Measurements were taken of the static flow-pressure relationship and linearity of the flow signal. Sensor airway resistance was calculated. Following recalibration of the host monitor, volumes ranging from 1 to 7 L were generated by a calibration syringe, and bias and precision of spirometric volume was determined. Where manual recalibration was not available, a conversion factor for volume measurement was determined. The influence of gas composition mixture and peak flow on the conversion factor was studied.
Results  Both the original and the remodelled sensor showed similar static flow–pressure relationships and linearity of the flow signal. Mean bias (%) of displayed values compared with the reference volume of 3, 5 and 7 L varied between −0.4% and +2.4%, and this was significantly smaller than that for 1 L (4.8% to +5.0%). Conversion factors for 3, 5 and 7 L were very similar (mean 6.00 ± 0.2, range 5.91–6.06) and were not significantly influenced by the gas mixture used. Increasing peak flow caused a small decrease in the conversion factor. Volume measurement error and conversion factors for inspiration and expiration were close to identity.
Conclusion  The combination of the host monitor with the remodelled flow sensor allowed accurate in vitro measurement of flows and volumes in a range expected during large animal anaesthesia.
Clinical relevance  This combination has potential as a reliable spirometric monitor for use during large animal anaesthesia.