首页 | 本学科首页   官方微博 | 高级检索  
     


Automatic flow system with voltammetric detection for diacetyl monitoring during brewing process
Authors:Rodrigues Pedro G  Rodrigues José A  Barros Aquiles A  Lapa Rui A S  Lima José L F C  Machado Cruz J M  Ferreira A A
Affiliation:CIQUP/Dep. Química da Faculdade de Ciências da Universidade do Porto, R. do Campo Alegre, 687, 4169-007, Portugal.
Abstract:Diacetyl can be determined by adsorptive stripping voltammetry after derivatization with o-phenylenediamine. The method may be applicable to the determination of diacetyl in different foods, being a good alternative to other analytical methods. In this work an on-line automated analytical system for diacetyl determination in beer is described. A hanging mercury drop electrode voltammetric flow detector was used, and the analyte was determined without the traditional deoxygenation procedure. The method was successfully applied to the determination of diacetyl during beer fermentation and in the final product. The automation strategy used was based on a flow network similar to those used in multicommutated flow systems, with a pervaporation unit used for diacetyl separation. The developed system was tested in real conditions in the monitoring of brewing processes. The results obtained were similar to those obtained with the usual GC-ECD methodology in the 5-600 ppb range. The analytical rate of the developed method is about 12 determinations/h.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号