首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of alpha-amanitin in serum and liver by multistage linear ion trap mass spectrometry
Authors:Filigenzi Michael S  Poppenga Robert H  Tiwary Asheesh K  Puschner Birgit
Institution:California Animal Health and Food Safety Laboratory System, Toxicology Laboratory, University of California, Davis, California 95616, USA. msfiligenzi@ucdavis.edu
Abstract:This paper describes a rapid LC-MS/MS/MS method for the analysis of alpha-amanitin in serum and liver. Serum was initially prepared by precipitation of proteins with acetonitrile and subsequent removal of acetonitrile with methylene chloride. Liver was prepared by homogenization with aqueous acetonitrile and subsequent removal of acetonitrile using methylene chloride. For both matrices, the aqueous phase was then extracted using mixed-mode C18/cation exchange SPE cartridges and analyzed on a linear ion trap LC-MS system. Standards were prepared in extracts of control matrix. Seven replicate fortifications of serum at 0.001 mug/g (1 ng/g) of alpha-amanitin gave a mean recovery of 95% with 8.8% CV (relative standard deviation) and a calculated method detection limit of 0.26 ng/g. Seven replicates of control liver fortified at 1 ng/g gave a mean recovery of 98% with 17% CV and a calculated method detection limit of 0.50 ng/g. This is the first report of a positive mass spectrometric identification and quantitation of alpha-amanitin in serum and liver from suspect human and animal intoxications.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号