首页 | 本学科首页   官方微博 | 高级检索  
     检索      

使用Li-6400测定森林土壤呼吸应注意的几个问题
引用本文:王慧梅,祖元刚,王文杰.使用Li-6400测定森林土壤呼吸应注意的几个问题[J].林业研究,2005,16(2):132-136.
作者姓名:王慧梅  祖元刚  王文杰
作者单位:[1]KeyLaboratoryofForestPlantEcologytheMinistryofEducation,NortheastForestryUniversity,Harbin150040,P.R.China [2]HokkaidoUniversityForests,FSC,HokkaidoUniversitySapporo060-0809,Japan
基金项目:国家自然科学基金,国家重点基础研究发展计划(973计划),Sina-Japan cooperation project on larch forest study
摘    要:正确使用Li-6400进行森林土壤呼吸测定是保证测定精度的前提.本文以落叶松林为例,对Li-6400使用过程应该注意的问题进行了研究.结果表明:塑料圈至少应该测定前12小时进行设置,以减少土壤CO2涌出效应对呼吸的影响,而且塑料圈插入深度对测定结果影响很大:当塑料圈插入过浅时横向气体扩散及塑料圈不稳定而造成测定时CO2的再涌出效应使土壤呼吸测定值偏高;而当插入深度过深时,切断根系导致根系呼吸下降,最活跃的土壤表层呼吸受塑料圈阻隔导致的土壤微生物呼吸测定值下降,这些现象共同造成显著低估土壤呼吸测定值.此外,在晴天白天大部分时间内目标CO2]值设定为外界相应时段的CO2]平均值,可以保证测定误差小于5%,而在清晨和傍晚进行测定时,则应该及时调整目标值为外界CO2]保证测量准确性.这一测定原则在实际测定中具有较高的实用性.

关 键 词:土壤呼吸  塑料圈设置深度  CO2涌出效应  横向扩散作用  参数选择
收稿时间:12 March 2005

Notes on the forest soil respiration measurement by a Li-6400 system
Wang Hui-Mei,Zu Yuan-Gang,Wang Wen-Jie,Koike Takayoshi.Notes on the forest soil respiration measurement by a Li-6400 system[J].Journal of Forestry Research,2005,16(2):132-136.
Authors:Wang Hui-Mei  Zu Yuan-Gang  Wang Wen-Jie  Koike Takayoshi
Institution:(1) Key Laboratory of Forest Plant Ecology, the Ministry of Education, Northeast Forestry University, 150040 Harbin, P. R. China;(2) Hokkaido University Forests, FSC, Hokkaido University, 060-0809 Sapporo, Japan
Abstract:The correct method used in forest soil respiration measurement by Li-6400 is a premise of data quality control. According to the study in a larch plantation, collars should be inserted at 12 hours in advance to efficiently reduce the influence of CO2 spring-out. Moreover, collar insertion depth substantially affected soil respiration measurement, i.e. when collar was shallowly inserted into soil, transversal gas diffusion and the CO2 re-spring-out caused by unstable collars in the measurement could lead to overestimating soil respiration rate; however, when collar was deeply inserted into soil, root respiration decline caused by root-cut and the most active respiratory of the surface soil separated by the inserted collars could lead to underestimating soil respiration rate. Furthermore, an error less than 5% could be guaranteed in typical sunny day if the target CO2] was set to the mean value of ambient CO2] in most time of the day, but it should be carefully set in early morning and late afternoon according to changing ambient CO2]. This protocol of measurement is useful in real measurement.
Keywords:Li-6400  Li-6400  Soil respiration  Collar insertion depth  CO2 spring-out effect  Gas transversal diffusion  Factory parameter selection
本文献已被 CNKI 维普 万方数据 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号