首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Comparison of four methods for digesting food samples for determination of trace levels of cadmium by flameless atomic absorption spectrophotometry
Authors:M T Cabanis  G Cassanas  J C Cabanis  S Brun
Institution:Laboratory of Analytical Chemistry and Industrial Physics, Faculty of Pharmacy, Montpellier, France.
Abstract:We compared 4 digestion procedures, namely, sulfuric-nitric acid in an open flask, nitric acid under pressure, sulfuric-nitric acid with refluxing, and nitric-hydrochloric-peroxide with refluxing, for the determination of cadmium by flameless atomic absorption spectrophotometry in 3 foodstuffs: rice, beef, and cream cheese. The foodstuffs were homogenized and divided into several batches for analysis. The results were evaluated using a 2-way cross analysis of variance. The study revealed that the digestion procedure was a highly significant factor (P less than 10(-4] in the analysis of the 3 foods; whereas the nature of the foodstuffs was not significant for rice and meat and only slightly significant (P less than 10(-2] for cream cheese. When the foodstuffs were spiked with a known amount of cadmium, we observed a loss of the metal when the sulfuric-nitric acid procedure in an open flask and the nitric-hydrochloric-peroxide digestion procedure were used. Taken together, the results of the present study indicate that the choice of the reagents used for digestion of foodstuffs is a crucial factor for cadmium determination by flameless atomic absorption spectrophotometry.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏