Authors:Rachael Gregson  R Eddie Clutton
Institution:Royal (Dick) School of Veterinary Studies, Division of Veterinary Clinical Sciences, The University of Edinburgh, Equine Hospital, Easter Bush Veterinary Centre, Midlothian, United Kingdom.
Abstract:A 7-year-old gelded Irish sports horse weighing 650 kg was anesthetized on 2 consecutive days for lavage of a septic right radio-carpal joint. On both occasions the endotracheal tube connector, which had been bound in medical tape to produce an airtight seal, functioned as a unidirectional valve during mechanical ventilation, retarding expiration, imposing positive end expiratory pressure (PEEP), and probably continuous positive airway pressure (CPAP). The equipment dysfunction was not identified on either occasion despite close inspection prompted by progressive increases in airway pressure and thoracic distension. Whilst the PEEP and CPAP exerted unexpectedly modest cardiovascular effects and the horse recovered uneventfully on both occasions, the improvisation may have proven fatal in a higher-risk subject.