首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Application of manual and automated systems for purification of ochratoxin a and zearalenone in cereals with immunoaffinity columns.
Authors:Mari Eskola  Meri Kokkonen  Aldo Rizzo
Institution:National Veterinary and Food Research Institute (EELA), Department of Chemistry, P.O. Box 45 (H?meentie 57), FIN-00581 Helsinki, Finland. mari.eskola@eela.fi
Abstract:A manual vacuum manifold and an automated solid phase extraction (ASPEC) system were applied for purification of ochratoxin A and zearalenone in wheat, rye, barley, and oat samples with immunoaffinity columns followed by separation with a high-performance liquid chromatograph and fluorescence detection. The immunoaffinity columns for manual sample purification were purchased from a different manufacturer than were those for the automated system. The limit of detection (LOD) for the method for ochratoxin A with a vacuum manifold and ASPEC was 0.1 microg/kg. For the method for zearalenone, the LODs were 1.5 microg/kg with a vacuum manifold and 3 microg/kg with ASPEC. For the methods for ochratoxin A at spiking levels of 0.6 and 2.5 microg/kg, mean recoveries for different cereals varied from 68 to 106%. For the methods for zearalenone, mean recoveries varied from 78 to 117% at spiking levels of 9 and 25 microg/kg. The relative standard deviations of repeatability with various cereals employing both methods were 2-15 and 2-19% for ochratoxin A and zearalenone, respectively.