An automatic flow procedure for the determination of 3-hydroxybutyrate in animal serum and plasma |
| |
Authors: | Pires Cherrine K Martelli Patrícia B Reis Boaventura F Lima José L F C Saraiva M Lúcia M F S |
| |
Affiliation: | Centro de Energia Nuclear na Agricultura, Universidade de S?o Paulo, PO Box 96, 13400-970, Piracicaba-SP, Brazil. |
| |
Abstract: | An automatic flow procedure based on the multicommutation concept, comprising three-way solenoid valves, for the spectrophotometric determination of 3-hydroxybutyrate in animal serum and plasma is proposed. The 3-hydroxybutyrate was enzymatically converted to acetoacetate with the reduction of NAD+ to NADH monitored at 340 nm. It was possible to carry out up to 600 determinations without a significant decrease in the analytical signal, with 5 mg of 3-hydroxybutyrate dehydrogenase immobilized on porous silica beads and packed in a column. The system enabled 60 determinations/h of 3-hydroxybutyrate in the range of 10-150 mg L(-1), with a consumption of 0.9 mg of NAD+ and 200 microL of sample per determination. A detection limit of 2 mg L(-1) for both animal serum and plasma and coefficients of variation of 1.4% and 1.2% (n = 17), respectively, were determined. Animal serum and plasma samples were analyzed without previous treatment, the results of which agreed with those obtained using the conventional method (UV kit, Sigma). |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|