首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Implantable cardioverter-defibrillator in a German shepherd dog with ventricular arrhythmias
Authors:Romain Pariaut DVM  Carley Saelinger VMD  Patricia Queiroz-Williams DVM  MS  Keith N Strickland DVM  H Cecilia Marshall DVM
Institution:aDepartment of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803, USA;bVeterinary Specialty Services, Manchester, MO 63021, USA
Abstract:This report describes the use of an implantable cardioverter-defibrillator (ICD) in a young German shepherd dog afflicted with inherited ventricular arrhythmias. Proper generator and lead placement was necessary for successful termination of ventricular fibrillation during device testing at the time of implantation. The risks of inappropriate therapy triggered by sinus tachycardia and oversensing of the T wave were controlled by extensive programming of the device. Following spontaneous resolution of the arrhythmia and due to the development of sepsis associated with the device, the ICD was successfully removed.
Keywords:Defibrillation  Canine  Sudden death  Ventricular ectopy  Ventricular fibrillation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏