首页 | 本学科首页   官方微博 | 高级检索  
     检索      

拉曼检测系统中微量试样自动混匀控制装置设计与试验
引用本文:闫帅,李永玉,彭彦昆,韩东海,刘亚超.拉曼检测系统中微量试样自动混匀控制装置设计与试验[J].农业机械学报,2021,52(1):324-332.
作者姓名:闫帅  李永玉  彭彦昆  韩东海  刘亚超
作者单位:中国农业大学工学院,北京100083;中国农业大学食品科学与营养工程学院,北京100083
基金项目:国家重点研发计划项目(2016YFD0101205)
摘    要:以提高表面增强拉曼光谱重复性为目的,为保证表面增强金属纳米颗粒和分析物混合吸附的均一性,基于实验室自行搭建的拉曼点检测装置,设计了表面增强剂与微量液态样品自动混匀控制拉曼光谱检测硬件系统.基于NI LabVIEW软件开发工具,采用G语言编写了实时控制分析软件,实现了微量液态样品和表面增强剂自动混匀进样、光谱采集及数据处...

关 键 词:蜂蜜  兽药残留  微量液体混匀  自动装置  快速检测  拉曼光谱
收稿时间:2020/3/18 0:00:00

Design and Test of Automatic Mixing and Control Device for Micro Sample in Raman Detection System
YAN Shuai,LI Yongyu,PENG Yankun,HAN Donghai,LIU Yachao.Design and Test of Automatic Mixing and Control Device for Micro Sample in Raman Detection System[J].Transactions of the Chinese Society of Agricultural Machinery,2021,52(1):324-332.
Authors:YAN Shuai  LI Yongyu  PENG Yankun  HAN Donghai  LIU Yachao
Institution:China Agricultural University
Abstract:To improve the repeatability of surface enhanced Raman spectroscopy(SERS)and ensure the homogeneity of the mixed adsorption of surface enhanced metal nanoparticles and analytes,a hardware system for automatic mixing control of surface enhancer and micro liquid samples was developed to ensure the uniform mixing effect of each sample with the surface enhancer based on the self-built laboratory Raman point detection device.At the same time,the real-time control analysis software was compiled with G language to implement one-button operations of automatic mixing and sampling of trace liquid samples and surface enhancers,spectrum acquisition and data processing by the NI LabVIEW software development tool.Finally,the stability of Raman detection system with automatic mixing of surface enhancers and micro liquid samples was verified by testing nitrofurantoin veterinary drug residues in commercially available honey.Compared with Raman point detection system,the Raman spectrum detection system for automatic mixing of surface enhancer and micro liquid sample exhibited a more satisfactory result that the lowest variation coefficients of Raman characteristic peaks for nitrofurantoin and sulfamethoxypyridazine were 0.0322 and 0.0361,respectively.In addition,the linear regression models of nitrofurantoin and sulfamethoxypyridazine in honey were established.The optimal validation set determination coefficients were 0.9619 and 0.9790,and the root mean square errors were 0.6723 mg/kg and 0.5186 mg/kg,respectively.These findings demonstrated that the system studied could significantly enhance the stability of Raman spectra,and the obtained quantitative prediction models of veterinary drugs were accurate,and it can offer technical support for the rapid quantitative detection of liquid samples by surface enhanced Raman spectroscopy.
Keywords:honey  veterinary drug residue  micro liquid mixing  automatic device  rapid detection  Raman spectroscopy
本文献已被 维普 万方数据 等数据库收录!
点击此处可从《农业机械学报》浏览原始摘要信息