首页 | 本学科首页   官方微博 | 高级检索  
     


FTIR approaches for diuron determination in commercial pesticide formulations
Authors:Armenta Sergio  Quintás Guillermo  Morales Asunción  Garrigues Salvador  de la Guardia Miguel
Affiliation:Department of Analytical Chemistry, University of Valencia, Edifici Jeroni Mu?oz, 50 Dr. Moliner, 46100 BURJASSOT, Valencia, Spain.
Abstract:Two strategies have been developed for Diuron determination by FTIR spectrometry, an off-line extraction and stopped-flow determination and a fully mechanized procedure, based on the on-line extraction of Diuron and FIA-FTIR measurement of the extracts. The aforementioned procedures have been compared with a reference chromatographic method. The off-line FTIR spectra were obtained at a nominal resolution of 4 cm(-1) from 4000 to 900 cm(-1) by accumulating 25 scans. Diuron was determined using peak height measurements at 1582 cm(-1) corrected using a baseline defined between 1562 and 1614 cm(-1). The waste generation of the off-line procedure was 3.4 mL chloroform for each sample, and the method provided a LOD of 40 microg g(-1), corresponding to 0.8% (w/w) Diuron in the original sample. The fully mechanized FIA method provided a LOD of 35 microg g(-1), which corresponds to 0.7% (w/w) in the solid sample and a maximum sampling frequency of the whole procedure of 30 h(-1), with a waste generation of 9.3 mL per sample, taking into account the volume of CHCl(3) required for sample dissolution and that need as a carrier. All those methods consume less organic solvent than a HPLC method, which involves the use of 39 mL of acetonitrile per sample and a sampling frequency of 12 h(-1).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号