Doppler blood pressure measurement in pigs during anaesthesia
Authors:Gabrielle C. Musk  Renata S. Costa  Jonathan Tuke
Affiliation:1. School of Veterinary and Life Sciences, Murdoch University, Australia;2. School of Mathematical Sciences, University of Adelaide, South Australia
Abstract:To determine the accuracy of Doppler blood pressure measurement in anaesthetised pigs eight large white pigs (Sus scrofa) were anaesthetised with zolazepam and tiletamine as Zoletil 100 (4 mg/kg) and xylazine (2 mg/kg) by intramuscular injection followed by isoflurane in 100% oxygen. Blood pressure measurements were recorded using a Doppler probe on the radial artery and a catheter in the femoral artery connected to a transducer. The sample mean of the Doppler measurements were compared with the sample mean of the invasive systolic arterial blood pressure while accounting for the heart rate, end-tidal CO2 and temperature. The predicted error of the Doppler was greatest when pigs were hypotensive and normothermic: 11 mmHg (CI: 6–15). Doppler measurements should be interpreted with caution in anaesthetised pigs that are hypotensive and normothermic.
Keywords:Pig   Anaesthesia   Blood pressure