首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High-throughput assay of oxygen radical absorbance capacity (ORAC) using a multichannel liquid handling system coupled with a microplate fluorescence reader in 96-well format
Authors:Huang Dejian  Ou Boxin  Hampsch-Woodill Maureen  Flanagan Judith A  Prior Ronald L
Institution:Brunswick Laboratories, 6 Thatcher Lane, Wareham, Massachusetts 02571, USA.
Abstract:The oxygen radical absorbance capacity (ORAC) assay has been widely accepted as a standard tool to measure the antioxidant activity in the nutraceutical, pharmaceutical, and food industries. However, the ORAC assay has been criticized for a lack of accessibility due to the unavailability of the COBAS FARA II analyzer, an instrument discontinued by the manufacturer. In addition, the manual sample preparation is time-consuming and labor-intensive. The objective of this study was to develop a high-throughput instrument platform that can fully automate the ORAC assay procedure. The new instrument platform consists of a robotic eight-channel liquid handling system and a microplate fluorescence reader. By using the high-throughput platform, the efficiency of the assay is improved with at least a 10-fold increase in sample throughput over the current procedure. The mean of intra- and interday CVs was
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏