The determination of creatinine in blood and plasma of small animals using the dry chemical system Reflotron
Authors:G Kridde  J Hirschberger
Institution:I. Medizinischen Tierklinik, Ludwig-Maximilians-Universit?t München.
Abstract:The reagent carrier Reflotron-Creatinine was tested for its suitability for whole blood and plasma samples of dogs and cats. It showed a very good correlation between the results of Reflotron-Creatinine and the reference method. Using plasma the results of both methods were extremely close. Using whole blood with a hematocrit of up to 50% good results could be achieved. If the hematocrit exceeded 50% large deviations could result.