The effect of anesthetic drug choice on accuracy of high-definition oscillometry in laterally recumbent horses
Authors:Tanya Duke-Novakovski  Barbara Ambros  Cindy Feng  Anthony P Carr
Institution:1. Department of Small Animal Clinical Sciences, Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon, SK, Canada;2. School of Public Health, University of Saskatchewan, Saskatoon, SK, Canada
Abstract:

Objective

To determine the accuracy of high-definition oscillometry (HDO) for arterial pressure measurement during injectable or inhalation anesthesia in horses.

Study design

Prospective, clinical study.

Animals

Twenty-four horses anesthetized for procedures requiring lateral recumbency.

Methods

Horses were premedicated with xylazine, and anesthesia induced with diazepam–ketamine. Anesthesia was maintained with xylazine–ketamine–guaifenesin combination TripleDrip (TD; n = 12) or isoflurane (ISO; n = 12)]. HDO was used to obtain systolic (SAP), mean (MAP) and diastolic (DAP) arterial pressures, and heart rate (HR) using an 8-cm-wide cuff around the proximal tail. Invasive blood pressure (IBP), SAP, MAP, DAP and HR were recorded during HDO cycling. Bland–Altman analysis for repeated measures was used to compare HDO and IBP for all measurements. The generalized additive model was used to determine if means in the differences between HDO and IBP were similar between anesthetic protocols for all measurements.

Results

There were >110 paired samples for each variable. There was no effect of anesthetic choice on HDO performance, but more variability was present in TD compared with ISO. Skewed data required log-transformation for statistical comparison. Using raw data and standard Bland–Altman analysis, HDO overestimated SAP (TD, 3.8 ± 28.3 mmHg; ISO, 3.5 ± 13.6 mmHg), MAP (TD, 4.0 ± 23.3 mmHg; ISO, 6.3 ± 10.0 mmHg) and DAP (TD, 4.0 ± 21.2 mmHg; ISO, 7.8 ± 13.6 mmHg). In TD, 26–40% HDO measurements were within 10 mmHg of IBP, compared with 60–74% in ISO. Differences between HDO and IBP for all measurements were similar between anesthetic protocols. The numerical difference between IBP and HDO measurements for SAP, MAP and DAP significantly decreased as cuff width:tail girth ratio increased toward 40%.

Conclusion and clinical relevance

More variability in HDO occurred during TD. The cuff width:tail girth ratio is important for accuracy of HDO.
Keywords:blood pressure  horse  isoflurane  oscillometry  TripleDrip