首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Modification of a Nonrebreathing Circuit Adapter to Prevent Barotrauma in Anesthetized Patients
Authors:R M MCMURPHY DVM  Diplomate ACVA    D S HODGSON DVM Diplomate ACVA  P H CRIBB BSC  MRCVS  Diplomate ACVA
Institution:Department of Clinical Sciences, Kansas State University, College of Veterinary Medicine, Manhattan, KS;Department of Anesthesiology, Radiology, and Surgery, Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
Abstract:Barotrauma, pneumothorax, and pneumomediastinum occurred in two anesthetized cats in which the waste gas outlet of a nonrebreathing circuit was occluded. To prevent any similar cases of barotrauma, we have modified our nonrebreathing circuit adapters by inserting a 15 cm H2O PEEP valve into the gas pathway of the nonrebreathing circuit adapter. This PEEP valve prevents the circuit and airway pressures from exceeding 15 cm H2O if the pop-off valve of the nonrebreathing circuit adapter is inadvertently left closed.