首页 | 本学科首页   官方微博 | 高级检索  
     检索      

一种国产近红外仪分析油菜籽三种主要品质参数研究
引用本文:智文良,信晓阳,崔建民,胡胜武,张文,李培武.一种国产近红外仪分析油菜籽三种主要品质参数研究[J].中国油料作物学报,2012,34(3):305-310.
作者姓名:智文良  信晓阳  崔建民  胡胜武  张文  李培武
作者单位:1.西北农林科技大学农学院,陕西 杨凌,712100;
2.中国农业科学院油料作物研究所,湖北 武汉,430062
基金项目:现代农业产业技术体系建设专项,西北农林科技大学拔尖人才支持计划
摘    要:利用中国农业科学院油料作物研究所研制的NYDL-3000智能型多参数粮油品质速测仪分析了油菜籽含油量、硫甙和芥酸含量,测定结果与传统化学方法进行了比较,其中含油量结果与PC-120核磁共振仪和MATRIX-I傅立叶变换近红外光谱仪做验证。内部交叉验证结果显示:含油量、硫甙和芥酸含量的决定系数R2分别为0.954 3、0.985 3和0.982 3,均方差分别为0.0469 、0.419 8和0.160 4,T检验显示三种品质参数交叉验证集的近红外光谱法(NYDL-3000)预测值和化学值差异均不显著。外部检验结果表明:硫甙含量、含油量两种方法测定值的相关系数分别为0.837 6和0.951 7,平均绝对误差分别为7.54%和0.02%,均在平均绝对误差允许的范围之内;成对数据双尾t检验表明,近红外光谱法和化学法两者结果差异不显著。芥酸分析中,NYDL-3000将342份样品中3份低芥酸材料判定为高芥酸,近红外光谱法和化学法两者结果符合率达99.12%。含油量的测定结果显示:NYDL-3000和PC-120、MATRIX-I结果的相关系数分别为0.956 6和0.901 5,平均绝对误差分别为0.78%和3.06%。研究结果表明,利用NYDL-3000智能型多参数粮油品质速测仪所建立的品质模型可以满足油菜育种对早代材料芥酸、硫甙含量和含油量的快速测定要求。

关 键 词:NYDL-3000多参数品质速测仪  近红外光谱  油菜  芥酸  硫甙  含油量

Determination of three major quality parameters of rapeseed with near infrared analyzer NYDL -3000
ZHI Wen-liang , XIN Xiao-yang , CUI Jian-min , HU Sheng-wu , ZHANG Wen , LI Pei-wu.Determination of three major quality parameters of rapeseed with near infrared analyzer NYDL -3000[J].Chinese Journal of Oil Crop Sciences,2012,34(3):305-310.
Authors:ZHI Wen-liang  XIN Xiao-yang  CUI Jian-min  HU Sheng-wu  ZHANG Wen  LI Pei-wu
Institution:1.College of Agronomy,Northwest Agriculture﹠Forest University,Yangling 712100,China; 2.Oil Crops Research Institute of Chinese Academy of Agricultural Sciences,Wuhan 430062,China)
Abstract:A near infrared reflectance spectrometer(NIRS) NYDL-3000 was designed by Oil Crops Research Institute of Chinese Academy of Agricultural Sciences.This research compared NYDL-3000 to traditional chemical methods in measuring three quality parameters of rapeseed including oil content,total glucosinolates and erucic acid.Oil content from NYDL-3000 was also compared to those from nuclear magnetic resonance spectrometer PC-120 and MATRIX-I Bruker Fourier transform NIR spectrometer.Coefficient of cross validation of oil content,total glucosinolates and erucic acid were 0.954 3,0.985 3 and 0.982 3;and RMSECV(root mean square error of cross validation) were 0.046 9,0.419 8 and 0.160 4 between NIRS and chemical methods.No obvious difference was found in cross validation between the two methods.External validation revealed that coefficient of correlation between NIRS and chemical methods were 0.837 6 and 0.951 7 for total glucosinolates and oil content.The compliance rate of erucic acid was 99.12% between NIRS and chemical method.No significant difference was found between the results of NIRS and chemical methods after tests of cross validation and external validation.Compared to other two spectrometers(as PC-120 and MATRIX-I),oil content results of NYDL-3000 respectively had coefficients of 0.956 6 and 0.901 5,and the RMSECV of 0.78%and 3.06%.It could be concluded that the NIRS NYDL-3000 analyzer was suitable for oil content,glucosinolates and erucic acid measurments on rapeseed breeding lines.
Keywords:NYDL-3000  Near infrared reflectance spectroscopy(NIRS)  Rapeseed  Erucic acid  Total glucosinolates  Oil content
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《中国油料作物学报》浏览原始摘要信息
点击此处可从《中国油料作物学报》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号