首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of shock waveform configuration on the defibrillation capacity of implantable cardioverter defibrillators in dogs
Authors:Romain Pariaut  Carley Saelinger  Jorge Vila  William Deforge  Patricia Queiroz-Williams  Hugues Beaufrère  Michael Zimmerman  Katrin Saile  Caryn A Reynolds
Institution:1. Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA, USA;2. St. Jude Medical, Cardiac Rhythm Management Division, Sylmar, CA, USA
Abstract:BackgroundImplantable cardioverter defibrillators (ICD) are programmed to detect ventricular arrhythmias and terminate them by delivering an electrical shock. A defibrillation threshold (DFT) at least 10 J below the maximum device output is recommended for successful therapy. Shock waveform configuration is a programmable parameter used to achieve a low DFT. It is hypothesized that a fixed-pulse configuration results in lower defibrillation energy requirements than a fixed-tilt configuration.Animals10 mongrel dogs.Materials and methodsICD generator and transvenous lead were surgically implanted. Defibrillation threshold was determined using a protocol guided by the upper limit of vulnerability. Fixed-pulse and fixed-tilt (50%/50%) waveform configurations were tested in a random order. Plasma cardiac troponin I (cTnI) was measured for signs of myocardial injury.ResultsThe experiment was completed in 9 dogs. Overall mean DFT value was 424 ± 88 V (9.2 ± 3.9 J). Mean differences among voltage, energy and impedance at the DFT for fixed-pulse (422 ± 97 V, 9.1 ± 4.2 J, 62.6 ± 13.8 Ω) and fixed-tilt (426 ± 83 V, 9.3 ± 3.8 J, 62.8 ± 18.5 Ω) configurations were not statistically significant (All P > 0.21). Cardiac TnI concentration changed from 0.03 ng/mL (95% CI: 0.02–0.04) at baseline to 0.11 ng/mL (95 CI: 0.08–0.16) after DFT was obtained with the first waveform configuration and 0.19 ng/mL (95% CI: 0.13–0.28) at the end of the study period. There were no significant changes in heart rate, end-tidal CO2 and blood pressure over time (all P > 0.09).ConclusionThe tested ICD device and lead placement reliably produced acceptable DFT values, based on a 10-J safety margin below the maximum device output. A benefit of fixed-pulse configuration could not be demonstrated over the standard fixed-tilt waveform. Signs of acute myocardial damage from repeated high-voltage shocks and episodes of ventricular fibrillation seemed of limited clinical significance.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号