Comparison of an infrared anaesthetic agent analyser (Datex‐Ohmeda) with refractometry for measurement of isoflurane,sevoflurane and desflurane concentrations
Authors:Andrea S Rudolff  Yves PS Moens  Bernd Driessen  Tamas D Ambrisko
Institution:1. Anaesthesiology and perioperative Intensive‐Care Medicine, Department for Companion Animals and Horses, University of Veterinary Medicine, , Vienna, Austria;2. Department of Clinical Studies‐NBC, School of Veterinary Medicine, University of Pennsylvania, , Kennett Square, PA, USA
Abstract:ObjectiveTo assess agreement between infrared (IR) analysers and a refractometer for measurements of isoflurane, sevoflurane and desflurane concentrations and to demonstrate the effect of customized calibration of IR analysers.Study designIn vitro experiment.SubjectsSix IR anaesthetic monitors (Datex-Ohmeda) and a single portable refractometer (Riken).MethodsBoth devices were calibrated following the manufacturer’s recommendations. Gas samples were collected at common gas outlets of anaesthesia machines. A range of agent concentrations was produced by stepwise changes in dial settings: isoflurane (0–5% in 0.5% increments), sevoflurane (0–8% in 1% increments), or desflurane (0–18% in 2% increments). Oxygen flow was 2 L minute?1. The orders of testing IR analysers, agents and dial settings were randomized. Duplicate measurements were performed at each setting. The entire procedure was repeated 24 hours later. Bland–Altman analysis was performed. Measurements on day-1 were used to yield calibration equations (IR measurements as dependent and refractometry measurements as independent variables), which were used to modify the IR measurements on day-2.ResultsBias ± limits of agreement for isoflurane, sevoflurane and desflurane were 0.2 ± 0.3, 0.1 ± 0.4 and 0.7 ± 0.9 volume%, respectively. There were significant linear relationships between differences and means for all agents. The IR analysers became less accurate at higher gas concentrations. After customized calibration, the bias became almost zero and the limits of agreement became narrower.Conclusions and clinical relevanceIf similar IR analysers are used in research studies, they need to be calibrated against a reference method using the agent in question at multiple calibration points overlapping the range of interest.
Keywords:anaesthetics  calibration  infrared gas analysis  monitoring  refractometry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号