Hematologic evaluation of normal and anemic lambs with the Technicon H*1 using EDTA or heparin as anticoagulants |
| |
Authors: | Vatn Synnøve Framstad Tore Torsteinbø Wenche O. |
| |
Affiliation: | Department of Sheep and Goat Research, Norwegian School of Veterinary Medicine, Sandnes, Norway;Borregaard Industries Ltd, Sarpsborg, Norway (Framstad). |
| |
Abstract: | Abstract: A study was performed to evaluate blood from young lambs using the Technicon H*1 hematology analyzer, with emphasis on RBC parameters, comparison of tripotassium EDTA and heparin, and the effects of storage on heparinized blood. Blood samples from lambs 2 days to 18 weeks of age were analyzed within 6 hours, revealing a high precision, except for WBC counts in heparinized blood. The HCT values estimated by the H*1 correlated well (r2= .90) with those obtained by the microhematocrit method. Mean hematologic values obtained for heparinized blood differed by up to 4% from values obtained for blood collected into EDTA. WBC counts decreased 8.4% in heparinized blood stored at 4°C for 1 day, but differences observed in RBC counts were ≤ 2%. Problems occurred when analyzing blood from young lambs with low hemoglobin values because the H*1 incorrectly counted microcytes with volumes of < 10 fL as platelets. When the necessary corrections were performed, the H*1 was useful for analyzing RBC parameters in lamb blood collected both into EDTA and into heparin. |
| |
Keywords: | Anemia EDTA hematology heparin lambs ovine Technicon H*1 |
本文献已被 PubMed 等数据库收录! |
|