In vitro evaluation of screws and suture anchors in metaphyseal bone of the canine tibia
Authors:Robb Julie L  Cook James L  Carson William
Institution:Comparative Orthopaedic Laboratory, University of Missouri-Columbia, Columbia, MO 65211, USA.
Abstract:OBJECTIVE: To compare ease of insertion, load to failure, and mode of failure of cortical and cancellous screws, BoneBiter, IMEX, and TwinFix suture anchors in canine metaphyseal tibial bone. STUDY DESIGN: Experimental biomechanical study. ANIMALS: Canine cadaveric tibias. METHODS: One investigator inserted all anchors and subjectively evaluated ease of placement. Anchor systems were loaded to failure along axis of insertion with audio-video recording to determine failure mode. RESULTS: BoneBiter was the most difficult anchor to insert successfully. Mean+/-SD loads to failure were cancellous screw (711+/-193 N), IMEX 4.7 mm 18 g wire (661+/-163 N), IMEX 4.0 mm 18 g wire (661+/-165 N), cortical screw (635+/-184 N), BoneBiter #5 Kevlar suture (393+/- 109 N), and TwinFix 5.0 mm #2 polyester (267+/-73 N). No significant differences were noted among the cortical screw, cancellous screw, IMEX 4.7 and 4.0 mm, all of which were significantly (P<.001) greater than BoneBiter and TwinFix . Failure modes were pullout of bone, suture-wire breakage, eyelet breakage, or no failure to 1000 N: screws (18,0,0,2), IMEX (18,1,1,0), BoneBiter (2,8,0,0), and TwinFix (0,10,0,0). CONCLUSIONS: Fixation devices were user friendly, with the exception of BoneBiter. Mode of failure is dependent on suture material and anchor design. Cortical and cancellous screws, and IMEX anchors with 18 g wire have significantly greater load to failure compared with BoneBiter and TwinFix suture anchors. CLINICAL RELEVANCE: Based on load to failure, ease of use, design characteristics, and cost, IMEX anchors may have advantages over other comparable soft tissue fixation devices.
Keywords:suture anchor  biomechanical testing  cancellous screw  cortical screw  BoneBiter?  TwinFix?  IMEX? anchors  Kevlar suture  Ethibond suture  bone  dog
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号