A comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats
Authors:Jepson Rosanne E  Hartley Vivien  Mendl Michael  Caney Sarah M E  Gould David J
Institution:Department of Clinical Veterinary Science, University of Bristol, Langford House, Bristol BS40 5DU, UK.
Abstract:Indirect blood pressure measurements were compared in 28 conscious cats using Doppler and oscillometric blood pressure-measuring devices. Ten cats were used to compare Doppler measurements between two examiners and 18 cats were used to compare Doppler and oscillometric measurements. The Doppler machine obtained systolic and diastolic blood pressure readings in 100% and 51% of attempts, respectively. With the oscillometric machine, systolic and diastolic blood pressure readings were obtained in 52% of the attempts. With the Doppler, measures of mean systolic blood pressure between two examiners were positively correlated, but there was no correlation for diastolic blood pressure measures. When comparing the results obtained by Doppler and oscillometric machines there was no significant difference between mean systolic blood pressure readings, but the oscillometric machine produced significantly higher estimates of diastolic blood pressure. In both cases, the standard deviations for the oscillometric machine were considerably larger than those for the Doppler machine. The first reading of systolic blood pressure obtained with the Doppler machine was an excellent predictor of the mean of five readings, but this was not so for the oscillometric machine. It took less than 5 min to obtain five readings in 37.5% of cases with the Doppler machine but this was true for only 5% of cases with the oscillometric machine. Two cats with ophthalmological lesions consistent with systemic hypertension were identified. In these two patients, systolic blood pressure measurements were between 200 and 225 mmHg when measured by Doppler, and between 140 and 150 mmHg when measured by the oscillometric machine. This suggests that a lower reference range for normal systolic blood pressure values should be used for the oscillometric device.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号