摘    要:全面对电镜真空系统的检查、分析,证实扩散泵向镜筒返油导致DXB2—12型速射电镜高真空故障,通过采用延时电路制止扩散泵油返入镜筒,排除了电镜高真空故障。

关 键 词:透射电镜,高真空,扩散泵,延时电路

ANALYSIS AND A REMOVING METHOD OF THE HIGH VACUVM FAULT IN A DXB2-12 TRANSMISSION ELECTRON MICROSCOPE
Yang Yusheng,Hou Yanping,Niu Jishan.ANALYSIS AND A REMOVING METHOD OF THE HIGH VACUVM FAULT IN A DXB2-12 TRANSMISSION ELECTRON MICROSCOPE[J].Journal of Shanxi Agricultural University,1996,16(1):97-99.
Authors:Yang Yusheng  Hou Yanping  Niu Jishan
Abstract:After examined and analysed the vacuum system of the electron microscope,it was discovered that the diffusion pump oie returning into the optical tube was the cause of the vacuum fault in DXB2-12 transmission electron microscope.To stop the diffusion oil returning into the optical tube the delay circuit was used.High vacuum fault was removed.
Keywords:Transmission elecfron microscope  High vacuum  Difusion pump  Delay circuit    