Assessment of stiffness and strength of 4 different implants available for equine fracture treatment: a study on a 20 degrees oblique long-bone fracture model using a bone substitute
Authors:Florin Marion  Arzdorf Michael  Linke Berend  Auer Joerg A
Institution:AO Research Institute, Davos, Switzerland.
Abstract:OBJECTIVE: To compare the mechanical properties of 4 stabilization methods for equine long-bone fractures: dynamic compression plate (DCP), limited contact-DCPlate (LC-DCP), locking compression plate (LCP), and the clamp-rod internal fixator (CRIF--formerly VetFix). STUDY DESIGN: In vitro mechanical study. SAMPLE POPULATION: Bone substitute material (24 tubes) was cut at 20 degrees to the long axis of the tube to simulate an oblique mid-shaft fracture. METHODS: Tubes were divided into 4 groups (n=6) and double plated in an orthogonal configuration, with 1 screw of 1 implant being inserted in lag fashion through the "fracture". Thus, the groups were: (1) 2 DCP implants (4.5, broad, 10 holes); (2) 2 LC-DCP implants (5.5, broad, 10 holes); (3) 2 LCP implants (4.5/5.0, broad, 10 holes) and 4 head locking screws/plate; and (4) 2 CRIF (4.5/5.0) and 10 clamps in alternating position left and right of the rod. All constructs were tested in 4-point bending with a quasi-static load until failure. The implant with the interfragmentary screw was always positioned on the tension side of the construct. Force, displacement, and angular displacement at the "fracture" line were determined. Construct stiffness under low and high loads, yield strength, ultimate strength, and maximum angular displacement were determined. RESULTS: None of the implants failed; the strength of the bone substitute was the limiting factor. At low loads, no differences in stiffness were found among groups, but LCP constructs were stiffer than other constructs under high loads (P=.004). Ultimate strength was lowest in the LCP group (P=.01), whereas yield strength was highest for LCP constructs (409 N m, P=.004). CRIF had the lowest yield strength (117 N m, P=.004); no differences in yield strength (250 N m) were found between DCP and LC-DCP constructs. Differences were found for maximum angular displacement at the "fracture" line, between groups: LPC
Keywords:biomechanical testing  oblique long-bone fracture model  dynamic compression plate (DCP)  limited contact–DCPlate  locking compression plate (LCP)  clamp-rod internal fixator (CRIF)  VetFix  bone substitute  equine  horse
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号