Evaluation the Clinitek status automated dipstick analysis device for semiquantitative testing of canine urine
Authors:Bauer N  Rettig S  Moritz A
Institution:aDepartment of Veterinary Clinical Sciences, Clinical Pathology and Clinical Pathophysiology, Justus-Liebig University Giessen, Frankfurterstrasse 129, 35392 Giessen, Germany
Abstract:The aim of this prospective study was to evaluate the Clinitek status™ analyser using Multistix10SG™/Microalbustix™ dipsticks (all: Siemens Dx) for canine urine (n = 101) compared to reference methods: visual reading (Combur9 dipstick, Roche), refractometry, microscopy, and quantitative protein/creatinine analysis (Pentra400, AxonLab). The automated analyses were done twice and visual tests were performed by two examiners.An excellent to good concordance was demonstrated between the first/second analysis with the Multistix10SG and the Combur9 dipstick, respectively with Cohen’s κ-values ranging from 0.776 to 1.000. Agreement between both dipsticks was good for glucose (κ = 0.753), blood (κ = 0.793), protein (κ = 0.788), and moderate for bilirubin (κ = 0.431) and ketones (κ = 0.540). In 6/101 specimens, false positive ketone reactions were obtained with the Multistix10SG™. Multistix10SG™ could not be used for determination of pyuria or specific gravity. Semiquantitative/quantitative protein results correlated well (ρ = 0.90) and creatinine measurements moderately (ρ = 0.76). Due to automated data transmission to the laboratory information system, the Clinitek status™ is of advantage in veterinary laboratories/clinics.
Keywords:Dog  Laboratory information system  Sediment  Sensitivity  specificity  Ketone
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号