Evaluation the Clinitek status automated dipstick analysis device for semiquantitative testing of canine urine |
| |
Authors: | Bauer N Rettig S Moritz A |
| |
Institution: | aDepartment of Veterinary Clinical Sciences, Clinical Pathology and Clinical Pathophysiology, Justus-Liebig University Giessen, Frankfurterstrasse 129, 35392 Giessen, Germany |
| |
Abstract: | The aim of this prospective study was to evaluate the Clinitek status™ analyser using Multistix10SG™/Microalbustix™ dipsticks (all: Siemens Dx) for canine urine (n = 101) compared to reference methods: visual reading (Combur9 dipstick, Roche), refractometry, microscopy, and quantitative protein/creatinine analysis (Pentra400, AxonLab). The automated analyses were done twice and visual tests were performed by two examiners.An excellent to good concordance was demonstrated between the first/second analysis with the Multistix10SG and the Combur9 dipstick, respectively with Cohen’s κ-values ranging from 0.776 to 1.000. Agreement between both dipsticks was good for glucose (κ = 0.753), blood (κ = 0.793), protein (κ = 0.788), and moderate for bilirubin (κ = 0.431) and ketones (κ = 0.540). In 6/101 specimens, false positive ketone reactions were obtained with the Multistix10SG™. Multistix10SG™ could not be used for determination of pyuria or specific gravity. Semiquantitative/quantitative protein results correlated well (ρ = 0.90) and creatinine measurements moderately (ρ = 0.76). Due to automated data transmission to the laboratory information system, the Clinitek status™ is of advantage in veterinary laboratories/clinics. |
| |
Keywords: | Dog Laboratory information system Sediment Sensitivity specificity Ketone |